Super clean workbench management

1. Objective: To maintain the biological cleanliness of the barrier system, to protect the safety of workers, to maintain animal welfare, to ensure the normal operation of the unit, and to minimize various variables that can affect scientific research results.
2. Principle: The ultra-clean workbench in the barrier system can only be used for the operation of animal models, the daily operation of transgenic mice.
3. Use management:
(1) Each time the ultra-clean workbench is used, the experimenter should first turn on the UV lamp on the ultra-clean workbench and use it after 20 minutes of UV irradiation.
(2) Turn on the working power of the ultra-clean workbench, turn off the UV lamp, and spray the disinfection work surface with 75% alcohol or 0.5% peracetic acid.
(3) During the whole experiment, the experimenter should operate according to the aseptic procedure.
(4) After the end of the experiment, wipe the work surface with disinfectant, turn off the working power, and turn on the UV lamp again for 15 minutes.
(5) In case of failure of the unit, the animal laboratory should be notified immediately, and the professional should continue to use it after passing the inspection.
(6) The experimenter should pay attention to keep the room clean and tidy.
(7) The filter material of the ultra-clean workbench is replaced once every 2 years, and the replacement record is made.

Shanghai Sujing Industrial Co., Ltd.
  
If you have any needs, you can always pass with me, thank you very much for your support to Sujing, look forward to working with you.
  
have a pleasant workday!
  
Thankyou&Bestregard!
  