Constant temperature incubator use method

1. Turn the power switch to “l”, then the power indicator light is on, and there is a digital display on the temperature controller;

2. Temperature setting

a. When the required cooking temperature is the same as the set temperature, it is not set without principle, otherwise it needs to be reset. First press the function key "SET" of the temperature controller to enter the temperature setting state, the SV setting display flashes and flashes, then press the shift key "â—¢" with the plus key "â–³" or the minus key "" Confirm with the function key "SET".

b. If you do not need to set 3 7~C, the original setting 2 6.5~C, first press the function key “SET”, then press the shift key “◢”, move the cursor to the display ten digits, then press “Add” △", so that the ten digits are raised from "2" to "3". After the ten digits are set, move the cursor to set the digits and quantile numbers in turn, so that the set temperature is displayed as 3 7~C, according to the function. Confirm with the key "SET" and the temperature setting ends.
3. Upper limit tracking alarm setting

The product has been set to a high of 10C before leaving the factory, and generally should not be set. If you need to reset the function key “SET” for 5 seconds, the meter enters the upper limit tracking alarm setting state “ALl” and then press the shift key “◢” with the plus key “△” or the minus key “▽” operation, and finally press the function. Confirm with the key "SET". The tracking alarm setting ends.

4. Select the culture time according to your needs. After the culture is finished, turn the power switch to “0”. If you do not remove the item immediately, please do not open the door.

5. After the setting is completed, each data is saved for a long time. At this time, the incubator enters a warming state, and the heating indicator lights up. When the temperature inside the box is close to the set temperature, the heating indicator lights up and down, and the control is entered into a constant temperature state.

6. Open the inner and outer doors, put the items to be cultivated into the incubator, and close the inner and outer doors. If the inner and outer doors open for too long, the temperature inside the box will fluctuate.
this is normal phenomenon.

7. If you have high requirements on temperature control accuracy and fluctuation, you can use PID self-tuning control. When the temperature inside the box reaches the set temperature for the first time, press the function key “SET” for 5 seconds, the meter enters. Set the cycle status "ALl", continue to press SET" button to display "ATU", SV display "0 0"
0 0", then press the plus key "â–³" to make SV display "00 01", and finally press the function key "SET" to confirm. At this time, the auto-tuning indicator lights, and the temperature controller enters the PID self-tuning control.

8. Temperature display value correction

Since the product has been rigorously tested before leaving the factory, it is generally not necessary to make corrections. If the environment is not good when the product is used, the outside temperature is too low or too high.
It will cause the temperature display value to be in error with the actual temperature inside the box. If it exceeds the technical specifications, it can be corrected. Specific steps: Press the function key "SET" 5
Second, the meter enters the parameter setting cycle state "ALl", continue to press the function key "SET", make "display "SC" correction, and then press the shift key ""
With the addition of the "â–³" or minus "â–½" operation, the temperature correction can be performed. Finally, press the "SET" button to confirm that the temperature display value correction is completed.